ProLink
®
II Software for Micro Motion
®
Transmitters
Initial Transmitter Startup Procedures
8. Wait until the zero time has expired. At the end of this time:
•
If the
Calibration Failure
status light turns red, the zero procedure failed.
-
See your transmitter manual for troubleshooting procedures.
-
If desired, use the buttons in the dialog box to restore the previous zero value or the
zero value established during factory calibration.
zero value established during factory calibration.
Note: These two functions are not available on all transmitters.
•
If the
Calibration in Progress
status light returns to green and the
Calibration Failure
status light does not turn red, the zero procedure succeeded.
9. Click
Close
.