5  Confirm that there is nothing on the pan and press 
the PRINT key. The balance measures the zero 
point. Do not allow vibration or drafts to affect the 
balance. 
  The balance displays the calibration weight value. 
6 Place a calibration weight, of the weight value 
displayed, on the pan and press the PRINT key. The 
balance measures the calibration weight. Do not 
allow vibration or drafts to affect the balance. 
7  The balance displays  end . Remove the weight 
from the pan. 
8  If the “GLP output (info)” parameter, of the function 
table, is set to “1” or “2”, the balance displays    glp   
and outputs “Calibration Report” using the RS-232C 
interface or stores the data in memory. For details 
on the calibration report format, refer to “11-2 GLP 
Report”. 
9  The balance will automatically return to the weighing 
mode. 
10  Place the calibration weight on the pan and confirm 
that calibration was performed correctly. If not, 
check the ambient conditions such as drafts or 
vibration, and repeat steps 2 through 10.   
 
 
 
 
 
 
 
 
 
 
 
 
 
 