Appendix D     Register-Level Programming
PC-LPM-16/PnP User Manual
D-30
 National Instruments Corporation
Clearing the Analog Input Circuitry
Write to the A/D Clear Register to clear the analog input circuitry, 
which leaves the analog input circuitry in the following state:
Analog input error flag OVERFLOW is cleared.
Pending interrupt requests are cleared.
To empty the A/D FIFO before starting any A/D conversions, perform 
two 8-bit reads on the A/D FIFO Registers and ignore the data read. 
This operation guarantees that the A/D conversion results read from the 
A/D FIFO are from the initiated conversions rather than leftover results 
from previous conversions.
To clear the analog input circuitry and the A/D FIFO:
Write 0 to the A/D Clear Register (8-bit write).
Programming Multiple A/D Conversions on a Single Input Channel Using 
Counter 0
This manual refers to a sequence of timed A/D conversions as a data 
acquisition operation. Counter 0 of the MSM82C53 is used as the 
sample-interval counter. In a data acquisition operation, counter 0 
continuously generates the conversion pulses. The software keeps track 
of the number of conversions that have occurred and turns off counter 0 
Table D-3.     Bipolar Input Mode A/D Conversion Values 
Input Voltage
A/D Conversion Result
Range: -5 to +5 V
(13-bit values)
A/D Conversion Result
Divided by 2
(12-bit values)
(Decimal)
(Hex)
(Decimal)
(Hex)
-5.0
-4,096
F000
-2,048
F800
-2.5
-2,048
F800
-1,024
FC00
0
0
0000
0
0000
2.5
2,048
0800
1,024
0400
4.9976
4,095
0FFF
2,047
07FF
a.Book : l.Appendix D  Page 30  Wednesday, November 20, 1996  6:36 PM