Omega Engineering FLR1000 사용자 설명서

다운로드
페이지 29
pg. 6 of 29 
1.  General Considerations 
 
It is recommended that a safety shut-off valve be installed upstream of 
(before) the sensor.  
 
All wetted parts should be checked for compatibility with the gas to be 
used. If there are any incompatibilities eg. highly corrosive gas, then the 
unit may be damaged or fail prematurely. Such damage will not be 
repaired under warranty. 
 
Units should be installed in a clean, dry environment with an ambient 
temperature that is as stable as possible. Avoid areas with strong magnetic 
fields, strong air flows or excessive vibration. 
 
If the gas to be used may contain particles larger than 25 microns then a 
filter (25 microns or less) should be installed upstream of (before) the unit.  
 
 
2.  Mounting the Flow Sensor or Flow Meter. 
 
Caution: On high flow units (2.0-10.0 L/min, and above) 
DO NOT disassemble the shunt / mounting plate 
assembly or the unit will not operate correctly and the 
warranty will be void. 
 
 
The Omega Engineering FLR 1000 Series units have no particular 
installation requirements so may be mounted in any convenient position. 
For optimum results the unit should be mounted with the mounting feet 
horizontal on top of a flat horizontal surface. This is the same position that 
the unit is calibrated in at the factory. Please contact the Omega 
Engineering Service Department if calibration for mounting in another 
orientation is desired.  
 
It is recommended that units be fixed to a suitable substrate with #4 
screws using the two mounting holes provided. 
 
Mounting The Sensor 
(mounting hardware not included with sensor)