 Conductivity Sensor  
Part 3 –Maintenance & Troubleshooting 
 
O&M Manual  
 
- 9 -  
Part 3 – Sensor Mounting 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
3.1 
Cleaning the Sensor 
 
Keep  the  sensor  as  clean  as  possible  for  optimum  measurement 
accuracy.  Frequency  of  cleaning  depends  upon  the  process 
solution.  
 
Note:  Mechanical  cleaning  of  the  electrode  surfaces  may  harm 
measurement quality if not performed with care. Do NOT use 
wire  brushes,  sandpaper  and  the  like  to  clean  any 
conductivity electrode.  
 
Wipe the measuring end of the sensor with a clean with a clean soft 
cloth.  Then rinse with clean water (distilled or de-ionized if 
possible).  This should remove most contaminate buildup.
 
 
If  necessary,  soak  the  sensor  for  several  minutes  in  a  mild  soap 
solution.  Use a small, extra-soft bristle brush (such as a mushroom 
brush)  to  thoroughly  clean  the  electrode  surfaces.    If  surface 
deposits  are  not  completely  removed  after  performing  this  step,  a 
dilute  acid  may  be  used  to  dissolve  the  deposits.    Soak  for  a  few 
minutes,  and  then  rinse  the  sensor  thoroughly  with  clean  water 
(distilled or de-ionized if possible). 
 
   
 
Note:  DO NOT soak the sensor in dilute acid solution for more 
 
 
 than 5 minutes. 
 
WARNING:  ACIDS  ARE  HAZARDOUS.    Always  wear  eye  and 
skin  protection  when  handling.    Follow  all  Material  Safety  Data 
Sheet  recommendations.    A  hazardous  chemical  reaction  can  be 
created  when  certain  acids  come  in  contact  with  process 
chemicals.  Make this determination before cleaning with any acid, 
regardless  of  concentration.    DO  NOT  use  Hydrochloric  Acid  on 
any stainless steel portion of the sensor. 
 
 
After cleaning the sensor, check measurement and re-calibrate the 
sensor.