Applying Finger (or Toe) Sensors
Adult Finger Sensors
Any finger except thumb on patients more than 50kg.
•
•
A-Series: M1191A or M1191AL
•
Special Edition (SE) A-Series: M1191ANL
•
B-Series: M1191B or M1191BL
Pediatric Finger Sensors
Any finger except thumb for patients between 15kg and 50kg
•
•
A-Series: M1192A
•
Special Edition (SE) A-Series: M1192AN
Infant Finger (or Toe) Sensors
Any finger or toe (except thumb) for patients between 4kg and 15kg.
Finger or toe diameter should measure from 7-8mm (0.27-0.31”).
•
Finger or toe diameter should measure from 7-8mm (0.27-0.31”).
•
A-Series: M1195A, or
•
Special Edition (SE) A-Series: M1195AN
Step
Applying the Finger (or Toe) Sensor
1
Select appropriate sensor for patient size (defined above).
2
Position sensor over patient’s finger. If applying an adult finger
sensor, be sure the sensor cable is positioned ON TOP of the fin-
ger/hand as shown above.
sensor, be sure the sensor cable is positioned ON TOP of the fin-
ger/hand as shown above.
3
Patient fingertip should touch but not protrude from end of sensor.
If necessary, trim the fingernail to correctly position the sensor.
If necessary, trim the fingernail to correctly position the sensor.
4
If applying an adult finger sensor, secure the cable to BACK OF
HAND with M1627A wristband (supplied only with adult finger
sensors.)
HAND with M1627A wristband (supplied only with adult finger
sensors.)
5
Plug sensor into instrument (or into adapter cable if needed).
6
Inspect and change the sensor application site periodically.
adult application
shown