Chapter 2
79
Instrument Functions: A - L
Auto Couple
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
2.1.6
ADC Dither
Access the menu to set
ADC Dither
to
On
,
Off
, or
Auto
.
Key Path:
Auto Couple
Remote Command:
[:SENSe]:ADC:DITHer[:STATe] OFF|ON|AUTO
[:SENSe]:ADC:DITHer[:STATe]?
Example:
:ADC:DITH OFF
2.1.6.1
Auto
Turns the ADC dither to automatic. It then chooses on or off according to which is most
likely to be the best selection, based on the other analyzer settings such as span, resolution
BW and sweep type (FFT or swept).
Key Path:
Auto Couple
State Saved:
Saved in instrument state
Factory Preset: Auto