Chapter 6 Ventilator Alarms
113
Alarm
Priority
Audible
Visual (Alarm 
Indicator Button 
and Display)
Device 
Action
User/Clinician Action
Low Inspiratory 
Pressure
High
• • •   • •
Red flashing button; 
“Low Inspiratory 
Pressure” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press 
Left (Reset) button to reset 
alarm. Check patient circuit 
for leaks or disconnects. If 
using an Active PAP or Active 
Flow circuit, check proximal 
lines to ensure they are 
not pinched or have water 
condensation (refer to the 
Troubleshooting section 
– tubing condensation). If  
alarm continues, contact your 
health care professional.
Clinician: Verify patient 
status. If problem continues, 
have device serviced.
High Minute 
Ventilation
High
• • •   • •
Red flashing button; 
“High Minute 
Ventilation” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press Left 
(Reset) button to reset alarm. 
Continue using device. If 
alarm continues, contact your 
health care professional.
Clinician: Verify patient 
status.
Low Minute 
Ventilation
High
• • •   • •
Red flashing button; 
“Low Minute 
Ventilation” message
Operates
User: Press Alarm Indicator/
Audio Pause button to 
silence alarm. Or, press Left 
(Reset) button to reset alarm. 
Continue using device. If 
alarm continues, contact your 
health care professional.
Clinician: Verify patient 
status.