Chapter 6 Ventilator Alarms
113
Alarm
Priority
Audible
Visual (Alarm
Indicator Button
and Display)
Device
Action
User/Clinician Action
Low Inspiratory
Pressure
Pressure
High
• • • • •
Red flashing button;
“Low Inspiratory
Pressure” message
“Low Inspiratory
Pressure” message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. Check patient circuit
for leaks or disconnects. If
using an Active PAP or Active
Flow circuit, check proximal
lines to ensure they are
not pinched or have water
condensation (refer to the
Troubleshooting section
– tubing condensation). If
alarm continues, contact your
health care professional.
Audio Pause button to
silence alarm. Or, press
Left (Reset) button to reset
alarm. Check patient circuit
for leaks or disconnects. If
using an Active PAP or Active
Flow circuit, check proximal
lines to ensure they are
not pinched or have water
condensation (refer to the
Troubleshooting section
– tubing condensation). If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status. If problem continues,
have device serviced.
status. If problem continues,
have device serviced.
High Minute
Ventilation
Ventilation
High
• • • • •
Red flashing button;
“High Minute
Ventilation” message
“High Minute
Ventilation” message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
status.
Low Minute
Ventilation
Ventilation
High
• • • • •
Red flashing button;
“Low Minute
Ventilation” message
“Low Minute
Ventilation” message
Operates
User: Press Alarm Indicator/
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Audio Pause button to
silence alarm. Or, press Left
(Reset) button to reset alarm.
Continue using device. If
alarm continues, contact your
health care professional.
Clinician: Verify patient
status.
status.