Chapter 7 Cleaning and Maintenance
125
5.  Place the circuit on a clean towel to dry. Do not wipe dry. The 
circuit must be completely dry before storing.
6.  Reassemble the circuit when dry. Store in a plastic bag or dust-
free area. 
7.  Inspect components for deterioration prior to use.
Cleaning the Exhalation Devices
Disassemble the exhalation device from the patient circuit. Follow 
the detailed cleaning instructions included with your exhalation 
device.
Replacing the Bacteria Filter (Optional)
To remove a bacteria filter and replace it with a new one:
1.  Disconnect the flexible tubing from the bacteria filter by gently 
pulling straight out from the filter. The proximal tubing (if using 
the Active PAP or Active Flow Exhalation Device) can remain 
connected to the ventilator.
2.  Remove the used bacteria filter from the ventilator by 
gently pulling straight out from the connector. Follow the 
manufacturer’s labeling and your institution’s infection control 
guidelines for disposal of the bacteria filter.
3.  Attach a new bacteria filter. The openings on either end of the 
bacteria filter are different sizes. Fit the larger opening onto the 
gas outlet port on the side of the ventilator as shown.
 
WARNING
Philips Respironics recommends 
that a main line outlet bacteria 
filter (Part Number 342077) be 
used whenever the device is 
used for invasive therapy or if 
the ventilator may be used on 
multiple patients.