Philips Trilogy 200 사용자 설명서

다운로드
페이지 181
Trilogy200 clinical manual
4
Personnel 
Qualifications
Trilogy200 is a restricted medical device designed for use by 
Respiratory Therapists or other trained and qualified caregivers under 
the supervision of a physician. 
The prescription and other device settings should only be changed on 
the order of the supervising physician.
The operator of the ventilator is responsible to read and understand 
this manual before use.
Modes of 
Ventilation
The device can provide therapies typically associated with both 
ventilator dependent and non-dependent patients. The mode of 
ventilation, circuit type, and alarm strategies should be chosen after a 
clinical evaluation of each patient’s needs.
C-Flex, Bi-Flex, and AVAPS are intended for use by adult patients.
SD Card 
Prescription 
Changes
When you change the device prescription, alarms, and other settings 
using the SD card, Trilogy200 requires that the caregiver review and 
verify the changes prior to the changes being used by the device. The 
caregiver or health care professional is responsible to ensure that the 
prescription settings are correct and compatible with the patient after 
using this feature. Installing the wrong prescription for a particular 
patient may result in improper therapy, lack of appropriate safety 
monitoring, and risk of death or injury to the patient.
Electrical 
Interference
This device is intended for use in the electromagnetic environment 
specified in Chapter 13 of this manual. The user of this device should 
make sure it is used in a compatible environment.
Portable and mobile RF communications equipment should be 
used no closer to any part of the device, including cables, than the 
recommended separation distance calculated using the information 
provided in Chapter 13 of this manual.
Battery Back-up 
Power
The internal battery is NOT intended to serve as a primary power 
source. It should only be used when other sources are not available or 
briefly when necessary; for example, when changing power sources.
The ventilator has a two-stage low battery alarm. The medium priority 
alarm indicates that approximately 20 minutes of operation remain, 
and the high priority alarm indicates that less than 10 minutes of 
operation remain. Actual run time may be more or less than this and 
varies with battery age, environmental conditions, and therapy.
Immediately seek an alternate power source when the “Low Battery” 
alarm appears. Complete power failure and loss of therapy is imminent.