Philips Trilogy 200 사용자 설명서

다운로드
페이지 181
Trilogy200 clinical manual
8
Maintenance
Follow the service recommendations provided in Chapter 7 of this 
manual.
Periodically inspect electrical cords, cables, and the detachable battery 
pack for damage or signs of wear. Discontinue use and replace if 
damaged.
Repairs and adjustments must be performed by Philips Respironics-
authorized service personnel only. Unauthorized service could cause 
death or injury, invalidate the warranty, or result in costly device 
damage.
Cleaning
(Refer to Chapter 7 
for detailed cleaning 
instructions.)
To avoid electrical shock, always unplug the power cord from the wall 
outlet before cleaning the ventilator. 
Do not immerse the device in any fluids or spray the device with water 
or cleaners. Clean the device with a cloth dampened with an approved 
cleaner.
If the device has been exposed to rain or dampness, dry the device 
including the area around the power cord connection with the power 
cord disconnected from the device before applying AC power.
  
Cautions
A caution indicates the possibility of damage to the device.
Storage
The internal and detachable batteries will self-discharge in storage. If it 
is desired to keep the batteries fully charged (for example, as a back-
up ventilator), plug the device into AC power for about eight hours 
every 16 days. Alternatively, the ventilator may be left continuously 
connected to AC power without battery degradation.
Allowing the batteries to fully discharge will not harm the batteries or 
lose device settings, but may require a longer battery charge time prior 
to use.