Philips Trilogy 200 사용자 설명서

다운로드
페이지 181
Trilogy200 clinical manual
24
Timed (T) Mode
In Timed (T) mode, the device delivers bi-level pressure support.  
This mode delivers Mandatory breaths only. A Mandatory breath 
is delivered according to the prescribed BPM and inspiratory time 
settings. This also means that the ventilator will not respond to 
patient effort. In this mode, an IPAP is delivered during inhalation and 
a lower EPAP is delivered during exhalation. The following illustration 
describes these concepts.
IPAP = 26 
cm H
2
O
EPAP = 6 
cm H
2
O
Rate = 10  BPM
PS = IPAP-EPAP = 20 
cm H
2
O
seconds
seconds
seconds
seconds
2
2
2
2
6
10
20
26
      = Mandatory Breath
2
Enter T mode
10 
15 
20 
25 
PRESSURE
TIME
T Mode