Philips Trilogy 200 사용자 설명서

다운로드
페이지 181
Chapter 3 Modes, Features, and Alarms
39
Spontaneous Expiratory Threshold (Inspiration to Expiration):
A second method used to cycle to expiration during spontaneous 
breathing in all modes except T, CV, AC, and SIMV, is called 
Spontaneous Expiratory Threshold (SET).  The SET rises in proportion 
to the inspiratory flow rate on each breath.  When the SET and actual 
patient flow value are equal, the unit cycles to expiration.
Pressure
Flow
IPAP
EPAP
Spontaneous
Expiratory
Threshold
Flow Reversal (Inspiration to Expiration):
As flow begins to decrease during inspiration, a flow reversal can 
occur due to a large leak around the mask or because the patient’s 
mouth is open. When the device senses this flow reversal, it 
automatically cycles to expiration.
Maximum IPAP Time (Inspiration to Expiration):
The maximum inspiratory time is determined by the adjustment of 
the Inspiratory time setting. A maximum IPAP time of 3.0 seconds 
acts as a safety mechanism to limit the time spent in inspiration 
during spontaneous breathing. For mandatory or assisted breaths, 
the maximum inspiratory time will equal the Inspiratory time setting 
up to 5.0 seconds.
Spontaneous Expiratory 
Threshold