Chapter 5 Viewing and Changing Settings
83
Pressure Control (PC) Mode
All of the settings available in S/T mode are available in PC mode, 
except for the Flow Cycle Sensitivity setting (when Flow Trigger is 
enabled). Please refer to the descriptions in the S and S/T Mode 
sections of this chapter for detailed information on the PC mode 
settings.
Pressure Control Synchronized Intermittent Mandatory Ventilation 
(PC-SIMV) Mode
The following settings, described in the S and S/T mode sections of 
this chapter, also are available in PC-SIMV mode:
 
Breath Rate
 
Inspiratory Time
 
Trigger Type
 
Flow Trigger Sensitivity (if Trigger Type is set to Flow 
Trigger)
 
Flow Cycle Sensitivity (if Trigger Type is set to Flow 
Trigger)
 
Rise Time
In addition to these, the following settings are also available in PC-
SIMV mode.
1.  Pressure
 
You can increase or decrease the Pressure setting from 4 to 50 
cm H
2
O in increments of 1. This is the pressure the ventilator will 
deliver during the inspiratory phase of a mandatory or assist 
breath.
2.  Pressure Support
 
You can increase or decrease the Pressure Support setting from 
0 to 30 cm H
2
O in increments of 1. This is the pressure support 
the ventilator will deliver during the inspiratory phase of a 
Spontaneous breath.
Note:  In PC-SIMV mode, you 
cannot set up Pressure Support 
for Mandatory and Assist 
breaths (Pressure - PEEP) greater 
than 30 cm H
2
O.
Note:  The Pressure Support and 
PEEP settings together cannot 
exceed 50 cm H
2
O.