Keep the following in mind when measuring resistance:
 
Because the meter’s test current flows through all
possible paths between the probe tips, the measured
value of a resistor in a circuit is often different from
the resistor’s rated value.
 
The test leads can add 0.1 
Ω
 to 0.2 
Ω
 of error to
resistance measurements. To test the leads, touch
the probe tips together and read the resistance of the
leads. If necessary, you can press 
D
 to
automatically subtract this value.
The resistance function can produce enough voltage to
forward-bias silicon diode or transistor junctions, causing
them to conduct. To avoid this, do not use the 30 M
Ω
 or
500 M
Ω
 ranges for in-circuit resistance measurements.
Testing for Continuity
Caution
To avoid possible damage to the meter or to
the equipment under test, disconnect circuit
power and discharge all high-voltage
capacitors before testing for continuity.
Continuity is the presence of a complete path for current
flow. The continuity test features a beeper that sounds if a
circuit is complete. The beeper allows you to perform
quick continuity tests without having to watch the display.
The continuity function detects intermittent opens and
shorts lasting as little as 1 millisecond (0.001 second).
These brief contacts cause the meter to emit a short beep.
To select continuity, turn the rotary switch to resistance
position, then press the blue button once. The continuity
symbol (
~
) appears in the display. Continuity uses
manual ranging only; autoranging is not available. Refer to
Figure 3-6 for continuity testing setup instructions.