Fluke 187 사용자 설명서

다운로드
페이지 93
Model 187 & 189
Users Manual
7-2
Physical Specifications
Display (LCD)
Digital: 50000/5000 counts primary display,
5000 counts secondary display;
updates 4/second.
Analog: 51 segments, updates 40/second.
Operating Temperature
 20 
°
C to + 55 
°
C
Storage Temperature
 40 
°
C to + 60 
°
C
Temperature Coefficient
0.05 x (specified accuracy) / 
°
C (<18 
°
C or >28 
°
C)
Relative Humidity
0 % to 90 % (0 
°
C to 35 
°
C)
0 % to 70 % (35 
°
C to 55 
°
C)
Altitude
Operating:  0-2000 meters per EN61010 CAT III, 1000 V; CAT IV, 600 V
                   0-3000 meters per EN61010 CAT II, 1000 V; EN61010 CAT III, 600 V;
                   CAT IV, 300 V
Storage: 10000 meters
Battery Type
4 AA Alkaline, NEDA 15A or LR6
Battery Life
72 Hours typical (with backlight off)
Shock Vibration
Per MIL-T-PRF 28800 for Class II instruments
Electromagnetic Compatibility
(EMC)
Susceptibility and Emissions: Commercial Limits per EN61326-1
Size
10.0 cm x 20.3 cm x 5.0 cm (3.94 in x 8.00 in x 1.97 in)
(Not Including Accessory Mount)
Weight
545 grams (1.2 lbs.)
Warranty
Lifetime
Calibration Interval
1 year