3M 245 사용자 설명서

다운로드
페이지 23
3M
TM 
Ranger
TM
 Blood/Fluid and Irrigation Warming Systems
English
8
Check the 3M
TM 
Ranger
TM
 system website to ensure you have the most recent version of this document. 
www.rangerfluidwarming.com reorder #202524A
Product Description
The Ranger blood/fluid, and irrigation fluid warming systems consist of a warming unit and a sterile 
disposable fluid  
warming set. 
ALERT
Alert indicator 
light
Slot for 
warming 
cassette
Alphanumeric 
display
Power switch
I.V. pole clamp
Bubble trap holder
m
odel
 245
ALERT
Alert
  
indicator 
light
Slot for 
warming 
cassette
Alphanumeric display
Power switch
I.V. pole clamp
m
odel
 247
r
anger
 w
arMIng
 u
nIt
The warming unit is a compact, lightweight, liquid-resistant device with a clamp located on the side 
for attachment to an I.V. pole. A carrying handle on the top of the unit makes it easy to transport. 
Located on the front panel you will find:
•  Alphanumeric display that indicates the heater temperature during normal operation. In an  
over-temperature condition, the display alternately flashes a temperature of 43°C for the Model 245 
(48°C for the Model 247) or higher and the word “HI.” An audible alert also sounds. In an under-
temperature condition, the display alternately flashes a temperature of 33°C or lower and the word 
“LO.”
•  Alert indicator light that comes on when either an over- or under-temperature condition occurs.