Philips V200 사용자 설명서

다운로드
페이지 328
Chapter 12
Technical Specifications
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-13
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The user 
of the ventilator should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test 
Level
Compliance 
Level
Electromagnetic Environment - 
Guidance
Electrostatic 
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
t±8kV air
Floors should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be 
at least 30%.
Electrical Fast 
Transient/burst
IEC 61000-4-4
±2kV for power 
supply lines.
±1kV for input/
output lines
±2kV for power 
supply lines.
±1kV for input/
output lines
Mains power quality should be 
that of a typical hospital 
environment
Surge
IEC 61000-4-5
±1kV differential 
mode
±2kV common 
mode
±1kV differential 
mode
±2kV common 
mode
Mains power quality should be 
that of a typical hospital 
environment
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines
IEC 61000-4-11
<5% U
T
 
(>95% dip in U
T
for 0.5 cycle 
40% U
T 
(60% dip in U
T
for 5 cycles 
70% U
T 
(30% dip in U
T
for 25 cycles 
5% U
T
 
(>95% dip in U
T
for 5 sec
<5% U
T
 
(>95% dip in U
T
for 0.5 cycle 
40% U
T 
(60% dip in U
T
for 5 cycles 
70% U
T 
(30% dip in U
T
for 25 cycles 
5% U
T
 
(>95% dip in U
T
for 5 sec
Mains power quality should be 
that of a typical hospital 
environment. If the user of the 
ventilator requires continued 
operation during power mains 
interruptions, it is recommended 
that the ventilator be powered 
from an external battery.
Power Frequency 
(50/60Hz) magnetic 
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic 
of a typical location in a typical 
hospital environment.
NOTE: U
T
 is the a.c. mains voltage prior to application of the test level.
Table 12-27: EMC Declaration - Immunity