Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ventilator is intended for use in the electromagnetic environment specified below. The user
of the ventilator should assure that it is used in such an environment.
of the ventilator should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Level
Compliance
Level
Level
Electromagnetic Environment -
Guidance
Guidance
Electrostatic
Discharge (ESD)
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
t±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical Fast
Transient/burst
Transient/burst
IEC 61000-4-4
±2kV for power
supply lines.
supply lines.
±1kV for input/
output lines
output lines
±2kV for power
supply lines.
supply lines.
±1kV for input/
output lines
output lines
Mains power quality should be
that of a typical hospital
environment
that of a typical hospital
environment
Surge
IEC 61000-4-5
±1kV differential
mode
mode
±2kV common
mode
mode
±1kV differential
mode
mode
±2kV common
mode
mode
Mains power quality should be
that of a typical hospital
environment
that of a typical hospital
environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be
that of a typical hospital
environment. If the user of the
ventilator requires continued
operation during power mains
interruptions, it is recommended
that the ventilator be powered
from an external battery.
that of a typical hospital
environment. If the user of the
ventilator requires continued
operation during power mains
interruptions, it is recommended
that the ventilator be powered
from an external battery.
Power Frequency
(50/60Hz) magnetic
field
(50/60Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
hospital environment.
should be at levels characteristic
of a typical location in a typical
hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Table 12-27: EMC Declaration - Immunity