Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-41
Chapter 13
Not used
N/A
N/A
N/A
Outputs example 
value.
Date
FEB 23 1997
N/A
N/A
N/A
12 character field, 
MMM DD YYYY
Mode setting
CMV
N/A
CMV
SIMV
CPAP
N/A
Assist/Control = 
CMV, Spont/Timed 
= SIMV, Spont = 
CPAP
Active Respiratory 
Rate setting
12.0
0.1 or 1
1.0 – 9.0
10 – 80
BPM
Outputs 
 
when mode is 
CPAP or Spont.
VCV Tidal Volume 
setting
0.50
0.01
0.05 – 2.50
L
VCV Peak Flow 
setting
5
 1
– 
140
LPM
Oxygen Concentration 
setting
100
1
21  –  100
%
Pressure Trigger 
setting
0.5
 0.1
0.1 
– 
20.0
cm 
H
2
O
Outputs value from 
active type (VCV or 
PCV). When active 
type is NPPV then 
outputs the VCV 
value.
PEEP or EPAP setting
0.0
 0.1
0.0 
– 
35.0
cm 
H
2
O
Uses PEEP setting 
from active 
ventilation type 
(VCV or PCV). 
When active type 
is NPPV uses 
EPAP setting.
VCV Insp. Hold 
Setting
0.0
 0.1
0.0 
– 
2.0
Sec
Not used
N/A
N/A
N/A
Outputs example 
value.
Not used
N/A
N/A
N/A
Outputs example 
value.
Not used
N/A
N/A
N/A
Outputs example 
value.
Not used
N/A
N/A
N/A
Outputs example 
value.
VCV Apnea Interval 
setting
10
 
1
10  –  60
sec
VCV Apnea Tidal 
Volume setting (same 
as Tidal Volume 
setting)
0.50
 
0.01
0.05 – 2.50
L
Outputs VCV Tidal 
Volume setting
Ventilation Report
Description
Example
Resolution
Range
Units
Comments
Table 13-4: Ventilation Report  (Sheet 2 of 10)