Philips V200 사용자 설명서
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-41
Chapter 13
Not used
N/A
N/A
N/A
Outputs example
value.
value.
Date
FEB 23 1997
N/A
N/A
N/A
12 character field,
MMM DD YYYY
MMM DD YYYY
Mode setting
CMV
N/A
CMV
SIMV
CPAP
N/A
Assist/Control =
CMV, Spont/Timed
= SIMV, Spont =
CPAP
CMV, Spont/Timed
= SIMV, Spont =
CPAP
Active Respiratory
Rate setting
Rate setting
12.0
0.1 or 1
1.0 – 9.0
10 – 80
BPM
Outputs
when mode is
CPAP or Spont.
CPAP or Spont.
VCV Tidal Volume
setting
setting
0.50
0.01
0.05 – 2.50
L
VCV Peak Flow
setting
setting
5
1
3
–
140
LPM
Oxygen Concentration
setting
setting
100
1
21 – 100
%
Pressure Trigger
setting
setting
0.5
0.1
0.1
–
20.0
cm
H
H
2
O
Outputs value from
active type (VCV or
PCV). When active
type is NPPV then
outputs the VCV
value.
active type (VCV or
PCV). When active
type is NPPV then
outputs the VCV
value.
PEEP or EPAP setting
0.0
0.1
0.0
–
35.0
cm
H
H
2
O
Uses PEEP setting
from active
ventilation type
(VCV or PCV).
When active type
is NPPV uses
EPAP setting.
from active
ventilation type
(VCV or PCV).
When active type
is NPPV uses
EPAP setting.
VCV Insp. Hold
Setting
Setting
0.0
0.1
0.0
–
2.0
Sec
Not used
N/A
N/A
N/A
Outputs example
value.
value.
Not used
N/A
N/A
N/A
Outputs example
value.
value.
Not used
N/A
N/A
N/A
Outputs example
value.
value.
Not used
N/A
N/A
N/A
Outputs example
value.
value.
VCV Apnea Interval
setting
setting
10
1
10 – 60
sec
VCV Apnea Tidal
Volume setting (same
as Tidal Volume
setting)
Volume setting (same
as Tidal Volume
setting)
0.50
0.01
0.05 – 2.50
L
Outputs VCV Tidal
Volume setting
Volume setting
Ventilation Report
Description
Example
Resolution
Range
Units
Comments
Table 13-4: Ventilation Report (Sheet 2 of 10)