Installation
Before installation, refer to the Safety Summary section at the front of this manual 
for power source, grounding, and other safety information.
Environment
Verify that you have the correct operating environment.
The AWG710&AWG710B Arbitrary Waveform Generator operates correctly in 
ambient temperatures from +10
°
 C to +40
°
 C and relative humidity from 20% to 
80%. If the instrument is stored long time at temperatures outside this range, wait 
for a while and then power on the instrument. For more operating environment 
information, refer to Appendix A:Specifications on page A-1.
Verify that there is nothing blocking the flow of air at the fan and air intake holes. 
The instrument exhausts air with the fan on its left side. Leave space at the sides of 
the instrument so that the instrument does not overheat. The following are the 
minimum space requirements for air flow around the instrument:
Rear 
7.5 cm (3 in)
Right and Left 
15.0 cm (6 in)
Top and Bottom
2 cm (0.8 in)
(The feet on the bottom of the instrument provide the required clearance when 
set on a flat surface.)
CAUTION. 
To prevent damage to the instrument, do not power on the instrument 
at temperatures outside the specified temperature range. 
NOTE. 
If you are installing the instrument in a rack, refer to the instruction sheet 
that comes with the rack–mounting kit for proper installation procedures.
NOTE. 
If the air flow is restricted and the internal temperature of the 
AWG710&AWG710B Arbitrary Waveform Generator exceeds the proper 
operating range, the instrument displays a message “Out of temperature limit” and 
temporarily shuts down to protect the internal modules from overheating. To 
prevent temporary shutdown of the 
AWG710&AWG710B Arbitrary Waveform 
Generator, do not restrict air flow through the chassis.
If the AWG710&AWG710B
 Arbitrary Waveform Generator
 shuts down 
unexpectedly, improve ventilation around the AWG710&AWG710B
 Arbitrary 
Waveform Generator
, and wait a few minutes to allow it to cool down; then 
switch the power on again.
You cannot power on the instrument when the ambient temperature exceeds the 
operation range. Wait until the instrument cools down, or the ambient temperature 
decreases to valid operating temperatures, before turning on the instrument again.