National Instruments NI 5422 사용자 설명서

다운로드
페이지 36
© National Instruments Corporation
17
NI 5422 Specifications
System Phase Noise and Jitter (10 MHz Carrier)
Sample Clock 
Source
System Phase Noise 
Density
(dBc/Hz) Offset
System Output Jitter
(Integrated from 
100 Hz to 100 kHz)
1. High-
Resolution 
specifications 
vary with 
Sample Rate.
2. Values are 
typical.
3. PXI Star 
trigger 
specification is 
valid when the 
Sample Clock 
Source is 
locked to 
PXI_CLK10.
100 Hz
1 kHz
10 kHz
Divide-by-N
–110
–122
–138
1.5 ps rms
High-
Resolution
1
100 MS/s
–109
–120
–120
4.0 ps rms
High-
Resolution
1
200 MS/s
–108
–120
–122
4.2 ps rms
CLK IN
2
–116
–130
–143
1.1 ps rms
PXI Star 
Trigger
2,3
–111
–128
–136
2.1 ps rms
External 
Sample Clock 
Input Jitter 
Tolerance
Cycle-Cycle Jitter ±150 ps
Period Jitter ±1 ns
Sample Clock Exporting
Exported 
Sample Clock 
Destinations
1. PFI<0..1> (SMB front panel connectors)
2. DDC CLK OUT (DIGITAL DATA & CONTROL front 
panel connector)
3. PXI_Trig<0..6> (backplane connector) 
Exported Sample 
Clocks can be 
divided by integer 
K (1 
≤ K ≤ 
4,194,304).
Exported 
Sample Clock 
Destinations
Maximum 
Frequency
Jitter (Typical)
Duty Cycle
PFI<0..1>
200 MHz
PFI 0: 6 ps rms
PFI 1: 12 ps rms 
25% to 65%
DDC CLK 
OUT
200 MHz
60 ps rms
35% to 65%
PXI_Trig<0..6>
20 MHz
Table 2.   (Continued)
Specification
Value
Comments