SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING
28
SECTION 4—SAFETY INSPECTION/
TROUBLESHOOTING
TROUBLESHOOTING
NOTE: Every six months or as necessary have a qualified technician perform a thorough inspection
and servicing. Routine maintenance will extend the life and efficiency of your tub.
and servicing. Routine maintenance will extend the life and efficiency of your tub.
Safety Inspection Checklists
Initially
❑ Ensure that the maximum water supply temperature does not exceed 115° F.
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Verify the water flow temperature with a separate thermometer.
❑ Visually inspect for cracks in the tub surface.
❑ Ensure that the door handle locks the door closed.
❑ Ensure that the safety pin properly engages the door.
❑ Check that all labels are present and legible. Replace if necessary.
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Verify the water flow temperature with a separate thermometer.
❑ Visually inspect for cracks in the tub surface.
❑ Ensure that the door handle locks the door closed.
❑ Ensure that the safety pin properly engages the door.
❑ Check that all labels are present and legible. Replace if necessary.
Before Each Use
❑ Disinfect the tub.
❑ Ensure that the maximum water supply temperature does not exceed 115° F.
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Visually inspect for cracks in the tub surface.
❑ Ensure that the door handle locks the door closed.
❑ Ensure that the safety pin properly engages the door.
❑ Ensure that the maximum water supply temperature does not exceed 115° F.
❑ Ensure that the maximum water pressure does not exceed 65 PSI.
❑ Visually inspect for cracks in the tub surface.
❑ Ensure that the door handle locks the door closed.
❑ Ensure that the safety pin properly engages the door.
Every 80-150 Baths
❑ Perform the heavy duty cleaning procedure on the tub. Refer to warning on page 30.
Every Six Months
❑ Verify the water flow temperature with a separate thermometer.
❑ Check that all labels are present and legible. Replace if necessary.
❑ Clean RADA valve to remove lime deposits.
❑ Check that all labels are present and legible. Replace if necessary.
❑ Clean RADA valve to remove lime deposits.