Emerson NGA2000 사용자 설명서
6 - 12
NGA 2000
90003749(1) [NGA-e (Software CLD 2.3/MLT 3.1.X)] 04/98
Main Menu — Technical level configuration
↓↓
System set up
↓↓
Module binding
↓↓
In the menu "Module Binding" you can bind with the analyzer modules the following types
of micro processor controlled I/O modules connected to the platform:
of micro processor controlled I/O modules connected to the platform:
1) I/O module with three alarms
2) Autocal I/O
3) Syscal I/O
2) Autocal I/O
3) Syscal I/O
You can unbind all existing bindings between the analyzer module and the I/O modules if
you press the F4 key (UNBIND) or if you press the ENTER key or the
you press the F4 key (UNBIND) or if you press the ENTER key or the
→
→
-key in the line
"Unbind everything!".
Module binding:
1) Select the channel you want:
♦
Change to the "Main Menu" with the F1 key (HOME).
♦
With the F3 key (NEXT) in the "Main Menu" you can select the analyzer module you
want to bind with an I/O module:
The channel selected will appear in the tag (left above in the headline of the display).
want to bind with an I/O module:
The channel selected will appear in the tag (left above in the headline of the display).
♦
Go back to the menu "Module Binding" via the menus "Technical level configuration"
and "System set up":
The tag of the analyzer module selected will appear in the line "Analyzer module
selected".
and "System set up":
The tag of the analyzer module selected will appear in the line "Analyzer module
selected".
2) Select the I/O module you want:
♦
Press the ENTER key or the
→
→
-key in the line "Select modules..." to change to the
submenu "Select I/O modules":
HOME
ESCAPE
BIND
UNBIND
INFO
CLD
F1
F2
F3
F4
F5
95.0 ppm
Module Binding
Select modules...
Proposed bind:
NONE
View bindings...
Bind selections!
Unbind everything!
Unbind everything!
Analyzer module selected:
CLD
6.1.4 System set up
Module Binding