Tektronix 1760 사용자 설명서

다운로드
페이지 180
Functional Overview
1760 Series Component Waveform/Vector Monitor User Manual
3-6
Overlay
OVERLAY
superimposes the selected input signals. In
OVER-
LAY
mode, the
LINE/FIELD
button remains a four-way toggle,
providing one line, two line, one field, and two field dis-
plays.
Sweep
Sweep buttons are used to select the waveform sweep rate.
LINE/FIELD
toggles through four sweep rate selections: one
line, two line, one field, and two field. In
PARADE
mode, the
LINE/FIELD
button becomes a two-way switch, toggling be-
tween line and field.
The sweep rate is displayed in the upper right corner of the
CRT (for field-rate sweeps, 1 F or 2 F is displayed).
The
MAG
button is used with
LINE/FIELD
to provide horizontal
magnification of each rate as follows:
H One line magnified = 200 ns/division
H Two line magnified = 1 µs/division
H One field or two field magnified = approximately