Adam Equipment azplus 15 사용자 설명서
6.2 USING “2003” TO ENTER THE SERVICE PARAMETERS
After the jumper has been placed in position K6, apply power to the scale.
During the self-test press the [Tare] key.
The display will ask for a code number, “
Pn - - - -
“ on the Weight Display.
Enter the number 2003 when “
Pn - - - -
“ is displayed and then press [Tare].
The displays will show the first parameters, called “
F1
”. To select another parameter press
the [PLU4] key to advance through the parameters.
Press the [Tare] key to enter a parameter.
To exit a parameter press the [Zero] key.
The Weight Display window will show the word “
CAL
” and the Unit Price window will show
the parameter number.
When a parameter is entered by pressing the [Tare] key, the displays will guide you
through the parameter selected and the options available.
The parameters available are:
F1
CAL
To enter the Calibration
F2
RES
Resolution
selection
F3
CAPA
Select
capacity
F4
0.00 Point
Weight decimal point location
F5
Point 0.00 0.00
Unit Price/Price decimal point
F6
Init ZEro
Initial Zero Range
F7
rE ZEro
Re-Zero range
F8
SCSivE tArE
Successive Tare Enable
F9
A/D counts
Display the A/D counts
F10
roUnd
Rounding off the Total Price
© Adam Equipment Company 2007