Troubleshooting the Trainer
The following table is provided to help you troubleshoot potential problems.
BEHAVIOR
POSSIBLE CAUSE
RECOMMENDED ACTION
Trainer says ANALYZING
before pads are applied to
manikin.
before pads are applied to
manikin.
• User touched the adhesive gel
on the pads and activated the
sensing circuitry of the Trainer.
sensing circuitry of the Trainer.
• Liner is not properly separating
pads.
• Wires are twisted.
• Avoid touching the adhesive gel
on the pads.
• Be sure the pads are pressed
firmly to the liner and only
touching each other through the
liner hole.
touching each other through the
liner hole.
• Untwist the wires.
Trainer does not say
ANALYZING when pads are
applied to the manikin
adapter.
ANALYZING when pads are
applied to the manikin
adapter.
• Pads placement is incorrect.
• Pads are old. Foil under adhesive
gel is severely cracked.
• Replaceable pads are not
properly connected to cartridge
cable.
cable.
• Manikin adapter is broken.
• User touched the adhesive gel
• User touched the adhesive gel
on the pad, so Trainer circuitry
did not sense that the pad had
been removed from the liner.
did not sense that the pad had
been removed from the liner.
• Pads wires are twisted.
• Check for proper pad placement.
Remove and reapply pads if
indicated.
indicated.
• Replace training pads.
• Check latch to be sure the
replaceable pads are correctly
attached to the cartridge cable.
attached to the cartridge cable.
• Replace manikin adapter.
• Avoid touching the adhesive gel
• Avoid touching the adhesive gel
on the pads.
• Untwist the pads wires.
Trainer says ANALYZING
while yellow liner is still
attached to pad.
while yellow liner is still
attached to pad.
Trainer cannot sense yellow liner.
Remove yellow liner.
NOTE: This cannot happen when
using the HS1 Defibrillator. The
defibrillator will not analyze heart
rhythm when the yellow liner is
attached to the pads. It senses that
the pads are not attached to a
patient and advises the user to
remove the yellow liner.
NOTE: This cannot happen when
using the HS1 Defibrillator. The
defibrillator will not analyze heart
rhythm when the yellow liner is
attached to the pads. It senses that
the pads are not attached to a
patient and advises the user to
remove the yellow liner.
The training pads cartridge
is difficult to install in or
release from the Trainer.
is difficult to install in or
release from the Trainer.
The resealable film cover may be
improperly aligned on the cartridge.
improperly aligned on the cartridge.
Make sure the resealable film cover
is lined up properly when repacking
the cartridge.
is lined up properly when repacking
the cartridge.
Trainer’s voice instructions
high-pitched and too rapid.
high-pitched and too rapid.
One of the batteries is installed
backwards.
backwards.
Be sure to install batteries
according to the diagram in the
Trainer’s battery compartment.
according to the diagram in the
Trainer’s battery compartment.