Troubleshooting the Trainer
The following table is provided to help you troubleshoot potential problems.
BEHAVIOR
POSSIBLE CAUSE
RECOMMENDED ACTION
Trainer says ANALYZING 
before pads are applied to 
manikin.
• User touched the adhesive gel 
on the pads and activated the 
sensing circuitry of the Trainer.
• Liner is not properly separating 
pads.
• Wires are twisted.
• Avoid touching the adhesive gel 
on the pads.
• Be sure the pads are pressed 
firmly to the liner and only 
touching each other through the 
liner hole.
• Untwist the wires.
Trainer does not say 
ANALYZING when pads are 
applied to the manikin 
adapter.
• Pads placement is incorrect.
• Pads are old. Foil under adhesive 
gel is severely cracked.
• Replaceable pads are not 
properly connected to cartridge 
cable.
• Manikin adapter is broken.
• User touched the adhesive gel 
on the pad, so Trainer circuitry 
did not sense that the pad had 
been removed from the liner.
• Pads wires are twisted.
• Check for proper pad placement. 
Remove and reapply pads if 
indicated.
• Replace training pads.
• Check latch to be sure the 
replaceable pads are correctly 
attached to the cartridge cable.
• Replace manikin adapter.
• Avoid touching the adhesive gel 
on the pads.
• Untwist the pads wires.
Trainer says ANALYZING 
while yellow liner is still 
attached to pad.
Trainer cannot sense yellow liner.
Remove yellow liner.
NOTE: This cannot happen when 
using the HS1 Defibrillator. The 
defibrillator will not analyze heart 
rhythm when the yellow liner is 
attached to the pads. It senses that 
the pads are not attached to a 
patient and advises the user to 
remove the yellow liner.
The training pads cartridge 
is difficult to install in or 
release from the Trainer.
The resealable film cover may be 
improperly aligned on the cartridge.
Make sure the resealable film cover 
is lined up properly when repacking 
the cartridge.
Trainer’s voice instructions 
high-pitched and too rapid.
One of the batteries is installed 
backwards.
Be sure to install batteries 
according to the diagram in the 
Trainer’s battery compartment.