Hanna Instruments hi 98143-01 사용자 설명서

다운로드
페이지 2
• Connect pH electrode, EC probe, power supply and
receiving devices to the HI98143 according to previous
instructions.
• Remove the protective cap from the pH electrode and
immerse it into the sample, at least 4 cm.
• When the matching pin is used dip it together
with the pH electrode, if matching pin is not in
use, connect the matching pin terminal to the
reference with a jumper wire.
• Immerse the EC probe in the sample.
• For continuous flow-thru measurements, the probes can
be easily installed in any standard ½” pipe tee. Use
Teflon tape between the probe and the pipe to ensure
a leak free joint. Take care not to overtighten it, as
excessive pressures can cause probe damage.
Note: to prevent damage to the electrode, disconnect the
pH electrode before turning the meter off.
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
For better accuracy, frequent calibration of the instrument is
recommended. The instrument should be recalibrated for
pH:
a) Whenever the electrode is replaced.
b) After testing aggressive chemicals.
c) At least once a month.
pH CALIBRATION
Pour small quantities of pH 7.0 (HI 7007)
and pH 4.0 (HI 7004) solution into two
clean beakers.
Calibration with pH 4.0 (HI 7004) solution is recommended
for measuring acidic samples. Use pH 10.0 (HI 7010)
solution if subsequent samples are alkaline.
• Connect a multimeter to the pH output terminals and
set it to Voltage readings.
Reference
Matching
Pin
• Turn the transmitter on.
• Rinse the electrode and immerse it in a pH
7.0 buffer with the grounding probe. Stir
gently and then wait for the reading to
stabilize.
• Adjust the pH Offset Calibration Trimmer with a small
screwdriver until the multimeter displays 500 mV.
• Rinse and immerse the pH electrode and
the grounding probe in pH 4.0 (or pH 10.0)
buffer and stir gently.
• Wait a couple of minutes and then adjust the pH Slope
Trimmer until the multimeter shows the value of the
second buffer expressed in mV (286mV for pH4 and
714mV for pH10 solution).
The pH calibration is now complete.
EC CALIBRATION
• Connect a multimeter to the EC output terminals and set
it to Voltage readings.
• Turn the transmitter on.
• Leave the HI3001 EC probe in air (dry probe).
• Adjust the EC Offset Calibration Trimmer with the cali-
bration screwdriver until the multimeter displays 0 mV.
• Pour a small quantity of 5.00 mS/cm calibration solu-
tion in a beaker.
• Immerse completely the tip of the EC probe into the
beaker with the conductivity solution.
• Tap the probe repeatedly on the bottom of the beaker
and stir it to ensure that no air bubbles are trapped
inside the sleeve.  In the case of in-line applications, to
better simulate the installation ambient,
the probe tip should be close to the
beaker bottom edge.
• Allow the reading to stabilize and then
adjust the EC Slope Calibration Trimmer
to read 500 mV on the multimeter LCD.
The EC calibration is now complete.
PH ELECTRODE MAINTENANCE
• Do not be alarmed if white crystals appear around the
electrode protective cap. This is normal with pH elec-
trodes and they dissolve when rinsed with water.
• When not in use, rinse the electrode with tap water to
minimize contamination and store it with a few drops of
storage (
HI 70300) or pH 7 (HI 7007) solution in the
protective cap.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR
STORAGE PURPOSES.
• If the electrode has been left dry, soak the tip in a
storage (
HI 70300) or pH 7 (HI 7007) solution for at
least one hour to reactivate it.
• To minimize clogging and provide longer life for the pH
electrode, it is recommended to clean it monthly. Im-
merse the tip of the electrode in 
HI 7061 for one hour
and then rinse it with tap water.
EC PROBE MAINTENANCE
• When calibration can no longer be achieved, remove
conductivity probe from the system for maintenance.
Rinse the probe with tap water. If a more thorough
cleaning is desired, remove the plastic protective sleeve
and clean the platinum sensors with a nonabrasive
cloth or HI 7061 cleaning solution.
Reinsert the sleeve in the same di-
rection as before, paying attention
the sleeve hole touches the threaded
edge.
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
ACCESSORIES
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
CE DECLARATION OF CONFORMITY
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residen-
tial areas could cause unacceptable interferences to radio and TV equip-
ment.  The glass bulb at the end of the electrode is sensitive to electro-
static discharges. Avoid touching this glass bulb all the time. The metal
band at the end of the EC electrode is sensitive to electrostatic dis-
charges. Avoid touching this metal band all the time. During operation,
ESD wrist straps should be
worn to avoid possible dam-
age to the electrode by
electrostatic discharges.
Any variation introduced by
the user to the supplied
equipment may degrade the
instrument’s EMC perfor-
mance.
To avoid electrical shock, do
not use this instrument when
voltages at the measurement
surface exceed 24 VAC or 60
VDC. To avoid damages or
burns, do not perform any
measurement in microwave
ovens.
ISTR98143-01R1  02/01
OUTPUT
OUTPUT
OUTPUT
OUTPUT
OUTPUT
NOTE:  The input is isolated from
output and power supply. The out-
puts aren’t isolated from power supply.
HI 1283
Stainless steel grounding probe with 2 m
(6.6’) cable
HI 1001
“flow-thru” double junction pH electrode with3
m (6.6’) cable and BNC connector
HI 3001
“flow-thru” 4-platinum ring EC probe with 3
m (6.6’) cable and temperature sensor
HI 7004M
pH 4.01 buffer solution, 230 mL bottle
HI 7007M
pH 7.01 buffer solution, 230 mL bottle
HI 7039M
5.00 mS/cm (EC) calibration solution, 230
mL bottle
HI 7061M
Electrode cleaning solution, 230 mL bottle
HI 710005 115 to 12 VDC power adapter
HI 710006 230 to 12 VDC power adapter
HI 77400P pH 4 & 7 solutions, 20 mL sachet (5 each)