National Instruments 3.21E+04 사용자 설명서

다운로드
페이지 181
Chapter 4     Signal Connections
AT-MIO/AI E Series User Manual
4-46
 National Instruments Corporation
General-Purpose Timing Signal Connections
The general-purpose timing signals are GPCTR0_SOURCE, 
GPCTR0_GATE, GPCTR0_OUT, GPCTR0_UP_DOWN, 
GPCTR1_SOURCE, GPCTR1_GATE, GPCTR1_OUT, 
GPCTR1_UP_DOWN, and FREQ_OUT.
GPCTR0_SOURCE Signal
Any PFI pin can externally input the GPCTR0_SOURCE signal, which 
is available as an output on the PFI8/GPCTR0_SOURCE pin.
As an input, the GPCTR0_SOURCE signal is configured in the 
edge-detection mode.  You can select any PFI pin as the source for 
GPCTR0_SOURCE and configure the polarity selection for either 
rising or falling edge.  
As an output, the GPCTR0_SOURCE signal reflects the actual clock 
connected to general-purpose counter 0.  This is true even if another PFI 
is externally inputting the source clock.  This output is set to tri-state at 
startup.
Figure 4-31 shows the timing requirements for the GPCTR0_SOURCE 
signal.
Figure 4-31.  GPCTR0_SOURCE Signal Timing
The maximum allowed frequency is 20 MHz, with a minimum pulse 
width of 23 ns high or low. There is no minimum frequency limitation.
The 20 MHz or 100 kHz timebase normally generates the 
GPCTR0_SOURCE signal unless you select some external source.
t
p
t
w
t
w
t
p
t
w
= 50 ns minimum
= 23 ns minimum