Calibration
 
 
C401/301/201/102/101 
 
 
 
23
3 CALIBRATION 
The C401/301/201/102/101 Colorimeter was calibrated and tested prior to 
leaving the factory. Therefore, it is possible for you to use the instrument 
directly out of the box. Re-calibration of the instrument is NOT 
recommended particularly in the case of chemical tests of very unstable 
and/or difficult to prepare calibration solutions. However, the meter allows 
you to calibrate in one point for each chemical test fitted in the instrument. 
It is recommended to calibrate the pH at 7.0 pH units using the supplied 
colorless 7.0 pH buffer. 
For each of the chemical tests fitted in the instrument, the Colorimeter 
requires only one standard solution for a complete calibration. During 
calibration, the instrument performs several system self-diagnostics. As 
such, several warning messages will be displayed. If the instrument 
detects an irregularity with either the detector or the lamp, a warning 
message will be displayed during or upon exiting the calibration mode. If 
this occurs, attempt to rectify the situation as indicated in Section 4 
Troubleshooting Guide.
 
In any event, the instrument will continue to read 
the concentration or pH, but with a decreased accuracy until the error is 
rectified.   
The meter has an auto exit feature whereby it will exits the calibration 
mode if no key is pressed within 10 minutes or if there is an error during 
calibration.  
A light shield cap is added in the package as an optional accessory to 
shield off stray light during calibration and measurements. 
3.1 
Calibration Procedure 
The standard calibration solutions are to be prepared in the ranges 
indicated in the specifications. Standard solutions should be measured 
accurately and qualified by a different method or another instrument of 
higher precision and accuracy.  
NOTE: Do not pour liquid directly into the sample well of the instrument. 
Always use a vial. The instrument will only accurately measure the sample 
when vials sealed with the black caps are used. The black cap serves as 
both seal and a light shield. 
NOTE: Do not attempt to clean the sample well. The optics may be 
damaged. 
NOTE: This instrument is designed to measure solutions contained in the 
custom made glass vials and therefore the use of other type of sample vials 
or sample cells may result in inaccurate measurements.