Calibration: pH
 
 
C401/301/201/102/101 
 
 
 
31
11.  Select another clean vial, un-screw the cap and fill it 
with 10 ml of the pH7.0 colorless buffer. (fill up to the 
line on the vial).   
Note: A proper calibration requires exactly 10 ml volume of pH7.0 
buffer. 
12.  Select a bottle of pH Indicator (Phenol Red) reagent. 
Unscrew the cap and holding the bottle vertically; gently 
squeeze 20 drops of reagent into the contents into the 
vial. Replace the cap onto the reagent bottle. Be careful 
not to over-tighten the cap.  
13.  Replace the cap onto the vial, ensuring it is screwed on 
finger tight.  Invert the vial several times to mix. 
14.  Gently place the vial into the sample well of the instrument, whilst 
again aligning the ▼ mark on the vial with the ▲ mark on the meter. 
15.  Wait for one minute and then press the READ/ENTER key within 3 
minutes from when the time the reagent was added to the vial. 
16.  The meter will display [--Rd--] whilst the calibration standard is being 
measured and then displays [SAVE] to prompt for calibration 
acceptance and saving of calibration data. 
    pH
Sample 
  7.0 pH
pH
READ
ENTER
CAL
 
17.  Press the READ/ENTER key. The meter will now accepts the 
calibration value and momentarily display [HoLD] and CAL 
annunciator before returning to the beginning of the calibration as in 
step 4. 
18.  To continue with another calibration, press the MODE key to select 
the desired calibration mode. 
19.  To exit the calibration mode, press the ZERO key to return to the 
measurement mode. 
Press ZERO to 
return to 
measurement mode
CAL
MODE
READ
ENTER
Press Mode to select 
another calibration test 
ZERO
CAL
 
20 Drops
  
  pH
Indicator
 Phenol
   Red
10 ml 
of pH 7.0