dhi use of molbox ohmic calibration cable pn 402150 사용자 설명서

다운로드
페이지 2
 
 
Use of molbox ohmic  
calibration cable 
P/N 402150 
Instruction Sheet 
This cable kit has two uses related to the molbox mass flow terminal. 
1)  This kit can be used to connect the following DHI molbloc simulators: 
P/N 401233 for 100 Ohm 
P/N 401232 for 107 Ohm 
P/N 401234 for 110 Ohm 
P/N 401235 for 115 Ohm 
to a precision resistance measurement standard for the purpose of measuring the resistance values of the 
precision resistors in the molbloc simulators. This is an important part of maintaining the measurement 
traceability of the molbox Ohmic Measurement System. 
2)  This kit can be used to connect (2) precision resistance references (decade boxes, for example) to the molbox 
mass flow terminal for the purpose of calibrating the molbox Ohmic Measurement System. This is in lieu of using 
molbloc simulators for the calibration of the molbox Ohmic Measurement System. 
CABLE USE FOR VERIFICATION OF MOBLOC SIMULATORS 
When using the cable to verify or measure the resistance values of the precision resistors within any of the DHI 
molbloc simulators, connect the 15-pin DSUB connector directly to the simulator connector.  There is no need to use 
the included gender changer.  The bananna plugs are then used to connect to the resistance measurement 
reference.  Use the 4-wire resistance measurement mode of the reference for a proper measurement. 
 
 
Figure 1 Cable use with DMM