Warnings and cautions
9
!
Caution: Unexpected system behavior.
Software viruses can harm computer programs. Cardiac Science 
recommends periodically performing a virus scan when 
CareCenter MD is not active. Virus scanning is especially important 
if CareCenter MD is connected to a network.
Electrical safety
The CareCenter MD is a medical electrical device. It meets all 
requirements for the safety of medical electrical devices, in 
particular the requirements for IEC/EN 60601-1, IEC/EN 
60601-1-2, and IEC/EN 60601-1-2-25.
The CareCenter MD is intended to be connected to a personal 
computer through a USB port. The personal computer and the 
peripheral devices (monitor, and printer) connected to the 
CareCenter MD should be certified to the appropriate safety 
standards for non-medical electrical equipment and must be 
located outside the patient environment.
If the personal computer or peripherals are located within the 
patient environment, an isolation transformer must be used to 
comply with IEC/EN 60601-1-1. 
For...
Isolation transformer status
CareCenter MD
Cardiac Science does not provide isolation 
transformers. The customer is responsible for 
providing appropriate isolation in the patient 
vicinity.
Quinton 9500 with 
CareCenter MD
Cardiac Science provides an isolation 
transformer.
Quinton 9550 with 
CareCenter MD
Cardiac Science provides an isolation 
transformer.