Warnings and cautions
9
!
Caution: Unexpected system behavior.
Software viruses can harm computer programs. Cardiac Science
recommends periodically performing a virus scan when
CareCenter MD is not active. Virus scanning is especially important
if CareCenter MD is connected to a network.
recommends periodically performing a virus scan when
CareCenter MD is not active. Virus scanning is especially important
if CareCenter MD is connected to a network.
Electrical safety
The CareCenter MD is a medical electrical device. It meets all
requirements for the safety of medical electrical devices, in
particular the requirements for IEC/EN 60601-1, IEC/EN
60601-1-2, and IEC/EN 60601-1-2-25.
requirements for the safety of medical electrical devices, in
particular the requirements for IEC/EN 60601-1, IEC/EN
60601-1-2, and IEC/EN 60601-1-2-25.
The CareCenter MD is intended to be connected to a personal
computer through a USB port. The personal computer and the
peripheral devices (monitor, and printer) connected to the
CareCenter MD should be certified to the appropriate safety
standards for non-medical electrical equipment and must be
located outside the patient environment.
computer through a USB port. The personal computer and the
peripheral devices (monitor, and printer) connected to the
CareCenter MD should be certified to the appropriate safety
standards for non-medical electrical equipment and must be
located outside the patient environment.
If the personal computer or peripherals are located within the
patient environment, an isolation transformer must be used to
comply with IEC/EN 60601-1-1.
patient environment, an isolation transformer must be used to
comply with IEC/EN 60601-1-1.
For...
Isolation transformer status
CareCenter MD
Cardiac Science does not provide isolation
transformers. The customer is responsible for
providing appropriate isolation in the patient
vicinity.
transformers. The customer is responsible for
providing appropriate isolation in the patient
vicinity.
Quinton 9500 with
CareCenter MD
CareCenter MD
Cardiac Science provides an isolation
transformer.
transformer.
Quinton 9550 with
CareCenter MD
CareCenter MD
Cardiac Science provides an isolation
transformer.