4.2 Digital mode
To operate the system:
1. Connect Wave Plus USB cable to the PC USB port. The power is taken directly from the USB
port;
The use of Personal Computer non-complying with IEC 60950 can cause electric shock
and damage the system. The use of Personal Computer non-complying with IEC 60950
is allowed using a medical class insulation transformer to power the PC.
and damage the system. The use of Personal Computer non-complying with IEC 60950
is allowed using a medical class insulation transformer to power the PC.
2. If required, connect the trigger cable to the external device, after checking trigger electric fea-
ture (voltage and timing) adopted by the system (read “Technical specifications” chapter);
3. Connect, if required, the Wave Plus power supply and switch on the system;