will  be  necessary  to  move  the  frequency  of  the  device  responsible  for  the 
interference  to  the  highest  available,  and  move  the  source  of  the  wireless  signal 
away from Wave Plus of at least 3 meters. 
 
 
 
How to use the device with disturbing electromagnetic fields 
 
If  strong  electromagnetic  fields  are  in  the  environment  where  Wave  Plus  works, 
some interference could occur during the data transfer via USB, causing a loss of 
EMG  data.  To  make  the  system  work  properly,  the  USB  cable  has  to  be  moved 
away from the source of noise.  
 
 
Use of the transportation Case 
 
If the system is provided with a transportation case (blue flight case), it is recom-
mended to always keep it with the downside facing downwards. The case can pro-
tect the system only from very light shocks, so it is NOT to be considered as an ab-
solute way of protecting it. Be sure to keep the key to the lock of the case in a safe 
place and away from children. 
 
 
How to remove the EMG probes positioned on the patient. 
 
The cables with clip connections used by standard wireless units for EMG (see pa-
ragraphs  3  and  3.1)  are  delicate  and  breakable  elements  of the  system,  not  to  be 
submitted to continuous strains and tractions which could partially or totally dam-
age them, with negative effects on the reliability of the EMG signal acquired. For 
this  reason,  we  suggest,  during  detachment  of  the  probes,  to  uncouple  the  clips 
from  the  pre-gelled  electrodes,  directly  handling  the  clips  and  without  stretching 
the cables, before proceeding to the complete removal of the EMG probe. 
 
Safety risk: 
 
The simultaneous connection of a patient to both a high frequency chirurgical de-
vice and an EMG system can cause burns where the electrical stimulation or biopo-
tential input takes place, and can possibly cause damages to the pre-amplifiers. 
 
More warnings: 
 
Do  not  use  Wave  Plus  system  in  an  environment  with  inflammable  anaesthetic 
mixture with air or oxygen or nitrogen protoxide.