VeriWave, Inc. WaveDeploy WiFi(WF1101)/Ethernet(EF1101) Users Guide
Page 24 of 26
AC Mains:
EN 55011 Class B limits for conducted emissions.
EN 61000-3-2 AC power line harmonic emissions.
6.9.2 Immunity, Enclosure, RF Electromagnetic Fields
No errors or other malfunctions when the instrument is subjected to a 27
MHz-1000 MHz, 10 V/m electromagnetic field 80% amplitude modulated
at 1kHz.
MHz-1000 MHz, 10 V/m electromagnetic field 80% amplitude modulated
at 1kHz.
6.9.3 Immunity, Enclosure, Electrostatic Discharge
Up to 8 kV with no change to control settings or impairment of normal
operation.
operation.
6.9.4 Immunity, Fast Transients, Electrical
No loss of stored data, change to control settings, degradation of
performance, or temporary loss of function will occur when the system is
subjected to the transients as describe below.
performance, or temporary loss of function will occur when the system is
subjected to the transients as describe below.
Port
Peak Voltage (kV) Tr/Th(ns)
Rep Freq. (kHz)
Signal & Control
0.5
5/50
5
AC Power
1
5/50
5
6.9.5 Immunity, AC Power Line Source Voltage Dips and Interruptions
Voltage Dips: 30% reduction / 10 mS, as per EN61000-4-11
Voltage Interruptions: >95% reduction / 5 Seconds, as per
EN6100-4-11
Voltage Interruptions: 100% reduction/5 cycle as per EN61326
Voltage Interruptions: >95% reduction / 5 Seconds, as per
EN6100-4-11
Voltage Interruptions: 100% reduction/5 cycle as per EN61326
6.9.6 Immunity, AC Power Line Transients
No loss of stored data, change to control settings, degradation of
performance, or temporary loss of function will occur when the system is
subjected to the transients as described below.
performance, or temporary loss of function will occur when the system is
subjected to the transients as described below.
Mode
Peak Voltage1 (kV)
Common
2
Differential
1
1 2/50 us Tr/Th voltage into open circuit, 8/20 us Tr/Th current into short
circuit, see specifications for additional detail.
circuit, see specifications for additional detail.