VeriWave, Inc.                                            WaveDeploy WiFi(WF1101)/Ethernet(EF1101) Users Guide 
 
 
Page 24 of 26 
 
 
 
AC Mains: 
 
EN 55011  Class B limits for conducted emissions. 
 
EN 61000-3-2  AC power line harmonic emissions. 
6.9.2   Immunity, Enclosure, RF Electromagnetic Fields 
No errors or other malfunctions when the instrument is subjected to a 27 
MHz-1000 MHz, 10 V/m electromagnetic field 80% amplitude modulated 
at 1kHz. 
6.9.3  Immunity, Enclosure, Electrostatic Discharge 
Up to 8 kV with no change to control settings or impairment of normal 
operation.  
6.9.4  Immunity, Fast Transients, Electrical 
No loss of stored data, change to control settings, degradation of 
performance, or temporary loss of function will occur when the system is 
subjected to the transients as describe below. 
Port 
Peak Voltage (kV)  Tr/Th(ns)  
Rep Freq. (kHz) 
Signal & Control 
0.5 
5/50 
AC Power 
5/50 
6.9.5  Immunity, AC Power Line Source Voltage Dips and Interruptions 
Voltage Dips: 30% reduction / 10 mS, as per EN61000-4-11 
Voltage Interruptions: >95% reduction / 5 Seconds, as per 
   EN6100-4-11 
Voltage Interruptions:  100% reduction/5 cycle as per EN61326 
6.9.6  Immunity, AC Power Line Transients 
No loss of stored data, change to control settings, degradation of 
performance, or temporary loss of function will occur when the system is 
subjected to the transients as described below. 
Mode 
   Peak Voltage1 (kV) 
Common 
 
Differential 
 
1 2/50 us Tr/Th voltage into open circuit, 8/20 us Tr/Th current into short 
circuit, see specifications for additional detail.