Guidance and manufacturer’s declaration – electromagnetic immunity 
 
The DIOMED DELTA 15/30 is intended for use in an electromagnetic environment specified below.  
The customer or the user of the DIOMED DELTA 15/30 should assure that it is used in such an 
environment.   
Immunity test 
IEC 60601 test level 
Compliance level 
Electromagnetic environment – 
guidance 
Electrostatic 
discharge (ESD) 
IEC 61000-4-2 
+/- 6 kV contact 
 
+/- 8 kV air 
+/- 6 kV contact 
 
+/- 8 kV air 
Floors should be wood, concrete or 
ceramic tile.  If floors are covered with 
synthetic material, the relative 
humidity should be at least 30%. 
Electrical fast 
transient / burst 
IEC 61000-4-4 
+/- 2 kV for power 
supply lines 
 
+/- 1 kV for 
input/output lines 
 
+/- 2 kV for power 
supply lines 
 
+/- 1 kV for 
input/output lines 
Mains power quality should be that of 
a typical commercial or hospital 
environment. 
Surge 
IEC 61000-4-5 
+/- 1 kV differential 
mode 
 
+/- 2 kV common 
mode 
 
+/- 1 kV 
differential mode 
 
+/- 2 kV common 
mode 
Mains power quality should be that of 
a typical commercial or hospital 
environment. 
Voltage Dips, short 
interruptions and 
voltage variations on 
power supply input 
lines 
IEC 61000-4-11 
 
<5% U
T
  
(>90% dip in U
T
for 0,5 cycle 
 
40 % U
T
  
(90% dip in U
T
for 5 cycles 
 
70 % U
T
  
(30 % dip in U
T
for 25 cycles 
 
<5% U
T
  
(>95% dip in U
T
for 5 sec 
 
<5% U
T
  
(>90% dip in U
T
for 0,5 cycle 
 
40 % U
T
  
(90% dip in U
T
for 5 cycles 
 
70 % U
T
  
(30 % dip in U
T
for 25 cycles 
 
<5% U
T
  
(>95% dip in U
T
for 5 sec 
Mains power quality should be that of 
a typical commercial or hospital 
environment.  If the user of the 
DIOMED DELTA 15/30 requires 
continued operation during mains 
interruptions, it is recommended that 
the DIOMED DELTA 15/30 be 
powered from an uninterruptible 
power supply or battery. 
 
Power frequency 
(50/60 Hz) 
IEC 61000-4-8 
 
3 A/m 
3 A/m 
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical 
commercial or hospital environment. 
 
NOTE    U
T
   is the a.c. mains voltage prior to application of the test level.