Guidance and manufacturer’s declaration – electromagnetic immunity
The DIOMED DELTA 15/30 is intended for use in an electromagnetic environment specified below.
The customer or the user of the DIOMED DELTA 15/30 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
+/- 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
transient / burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
supply lines
+/- 1 kV for
input/output lines
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
supply lines
+/- 1 kV for
input/output lines
Mains power quality should be that of
a typical commercial or hospital
environment.
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
IEC 61000-4-5
+/- 1 kV differential
mode
+/- 2 kV common
mode
mode
+/- 2 kV common
mode
+/- 1 kV
differential mode
+/- 2 kV common
mode
differential mode
+/- 2 kV common
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
a typical commercial or hospital
environment.
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% U
T
(>90% dip in U
T
)
for 0,5 cycle
40 % U
40 % U
T
(90% dip in U
T
)
for 5 cycles
70 % U
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5% U
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>90% dip in U
T
)
for 0,5 cycle
40 % U
40 % U
T
(90% dip in U
T
)
for 5 cycles
70 % U
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5% U
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
DIOMED DELTA 15/30 requires
continued operation during mains
interruptions, it is recommended that
the DIOMED DELTA 15/30 be
powered from an uninterruptible
power supply or battery.
a typical commercial or hospital
environment. If the user of the
DIOMED DELTA 15/30 requires
continued operation during mains
interruptions, it is recommended that
the DIOMED DELTA 15/30 be
powered from an uninterruptible
power supply or battery.
Power frequency
(50/60 Hz)
IEC 61000-4-8
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.