DELTA 15/30 Operator Manual 
Page 2.19 
Version B.2 
 
 
Recommended separation distances between portable and mobile RF communications and the 
DIOMED DELTA 15/30. 
The DIOMED DELTA 15/30 is intended for use in an electromagnetic environment in which 
radiated RF disturbances are controlled.  The customer or the user of the DIOMED DELTA 15/30 
can help prevent electromagnetic interference by maintaining distance between portable and 
mobile RF communications equipment (transmitters) and the DIOMED DELTA 15/30 as 
recommended below, according to the maximum output power of the communications equipment. 
 
Separation distance according to frequency of transmitter 
Rated maximum output power of 
transmitter 
150kHz to 80 MHz 
d = 1.2  √P 
80MHz to 800MHz 
d = 1.2  √P 
800 MHz to 2,5 GHz 
d = 2.3 √P 
0.01 
0.12 
0.12 
0.23 
0.1 
0.38 
0.38 
0.73 
1.2 
1.2 
2.3 
10 
3.8 
3.8 
7.3 
100 
12 
12 
23 
 
For transmitters rated at a maximum output power not listed above, the recommended separation 
distance d in metres (m) can be estimated using the equation applicable to the frequency of the 
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according 
to the transmitter manufacturer. 
 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected 
by absorption and reflection from structures, objects and people.