Appendix B – Calibration, Traceability, and Recalibration
Ophir Pyroelectric Heads
The Original area displays the original Measurement and Calibration
Factor. The Current wavelength is displayed beneath the Original area.
Factor. The Current wavelength is displayed beneath the Original area.
Note: A different wavelength can be selected in the Measurement Parameters
area.
2. Apply the known laser power.
3. Adjust the Calibration Factor until the Adjusted measurement equals the
power applied to the head.
4. Click Save to save the adjustment for the active channel.
Note: For metallic heads, when the calibration is changed at one laser
wavelength, the overall calibration of all other wavelengths changes
proportionately. For broadband heads, there is an option to adjust the
calibration factor for all wavelengths or only for a selected
wavelength.
proportionately. For broadband heads, there is an option to adjust the
calibration factor for all wavelengths or only for a selected
wavelength.