Appendix B – Calibration, Traceability, and Recalibration  
Ophir Pyroelectric Heads 
The Original area displays the original Measurement and Calibration 
Factor
. The Current wavelength is displayed beneath the Original area. 
Note: A different wavelength can be selected in the Measurement Parameters 
area. 
2.  Apply the known laser power. 
3.  Adjust the Calibration Factor until the Adjusted measurement equals the 
power applied to the head.  
4.  Click Save to save the adjustment for the active channel. 
Note: For metallic heads, when the calibration is changed at one laser 
wavelength, the overall calibration of all other wavelengths changes 
proportionately. For broadband heads, there is an option to adjust the 
calibration factor for all wavelengths or only for a selected 
wavelength. 