Event Types
Link Alarm
This alarm occurs when there is a problem linking the infant to a mother. For example, the infant may already be 
linked to another mother.
To respond to a Link Alarm
1.    Always follow your facility’s policies and procedure to ensure patient safety and secure the area.
2.    Click anywhere in the Link Alarm Message Box to access the Event Information window.
3.    Click Clear.
4.    If necessary, relink the mother and infant.
Check Alarm
No message has been received from the patient transmitter or the device in the patient room during the Inactivity 
Check-In period. The Check message will remain until an Alarm message is received, or the event is manually 
cleared at the Client computer(s).
To respond to a Check Alarm
1.    Always follow your facility’s policies and procedures to ensure patient safety and secure the area.
2.    Locate the device. Check Alarms must be reset at the transmitting device.
3.    Verify that no patient emergency exists.
4.    From the Client computer, click anywhere in the Inactivity Check-In Alarm Message Box to access the 
Event Information window.
5.    Select Clear to clear the Inactivity Check-In alarm.
N
OTE
You can click Silence to stop the alarm sound at the Central Server or Client 
computer. The alarm is silenced, at that computer only, for the configured length of time; 
however, the event still appears in the Event List. The next event automatically restarts the 
alarm sound.