Appendix: Electrical Safety 
LED Color 
State 
Action Recommended 
Red 
Charging 
Continue charging until the LED turns green. 
Yellow 
A charging circuit fault has 
occurred 
Undock and then re-dock the data receiver in the 
docking station.  If the yellow light persists 
contact technical support. 
Green 
Fully charged 
The data receiver is ready for use. 
Off 
x Docking station is not 
connected to an AC power 
source 
x Receiver is not fully connected 
to the docking station 
x Docking station or data 
receiver are in thermal 
shutdown 
x Connect the docking station to a source of 
AC power. 
 
x Undock and then re-dock the data receiver in 
the docking station. 
x Contact technical support for assistance. 
Activation Fixture 
 
The activation fixture turns the capsule on and off using strong magnets that interact 
with the capsule’s internal power switch.  
Warning 
Individuals with pacemakers should not come within one (1) foot of the SmartPill activation fixture. The 
fixture contains strong magnets that could interfere with pacemaker operation. 
Warning 
Do not store the SmartPill Activation Fixture in the same room with or a room adjacent to MRI equipment. 
The fixture contains strong magnets and could become a dangerous projectile. 
Caution 
Keep the SmartPill Activation Fixture more than two (2) feet from magnetic media and computer monitors. 
Caution 
Do not store unused capsules within one (1) foot of the SmartPill Activation Fixture. Stray magnetic fields 
from the activation fixture may activate the capsule. 
System Computer and MotiliGI Software 
MotiliGI software comes installed on the system computer. MotiliGI receives and 
processes downloaded data from the data receiver, stores test data, provides data 
analysis tools, and graphically displays test results. MotiliGI features algorithms that 
calculate GET, SBTT, CTT, WGTT, and motility indices of the antrum and