Appendix: Electrical Safety 
6WDUWLQJD7HVW
Preparing the System 
Perform these steps immediately before or after a patient’s arrival at the office. 
1. Turn on the system computer. 
2. Open MotiliGI: 
x Click  
on 
the 
desktop. 
x Select Start > All Programs > SmartPill > MotiliGI. 
3. If prompted, enter a username and password. Click OK.  The main startup screen 
appears. 
Performing the Test 
± Test Initiation Wizard 
Step 1: Connect the Data Receiver 
Caution 
Do not close MotiliGI while operating the Test Initiation Wizard because data can be lost.  Complete or 
cancel the Wizard first.  
Caution 
You may cancel the wizard any time. However, a capsule exposed to the pH calibration buffer must be 
ingested within 2 hours or discarded. 
1. Follow the prompts. Check the results: 
Component 
Success 
Troubleshooting 
Docking station’s 
LED 
Green 
x Red 
The data receiver may not be fully charged. See Chapter 4: 
Charging the Data Receiver. 
x Yellow 
Undock and re-dock the data receiver in the docking 
station. If the yellow light persists, contact SmartPill 
Technical Support. 
Data receiver 
display 
The “Looking For 
Capsule” message 
appears. 
x LCD is blank 
Turn on the data receiver by pressing the event button. 
System 
Computer 
x 2 beeps 
x These icons 
appear in the 
lower-right 
corner of the 
screen: 
x
 
x No beeps 
x Check the computer’s sound level. 
x Undock and re-dock the data receiver in the docking 
station. If there are still no beeps, contact Technical 
Support. 
x The “Receiver is Not Fully Charged” message appears. 
The charge is insufficient for a 5-day test.