- 5 - 
 
Located on the RX10K rear panel is the On/Off Power switch, a 
jack for 12VDC Power Input, a Fuse housing, a connector for 
attaching the Receiver Antenna, binding posts for the analog 
Voltage Output, and a Com (DB9) connector for the digital data 
signal.  The analog Voltage Output signal has a nominal range of 
±10VDC and a maximum range of ±12VDC.  The digital data 
signal is an RS-232 type signal for input to a PC “Com” port.  See 
Appendix A for the pin out and protocol.   
 
Power 
Input
Antenna
12VDC
1 Amp
Fuse
On
Power
Off
Com
Voltage
Output
+
Binsfeld Engineering Inc.   Maple City, MI    USA
      (+1) 231-334-4383     www.binsfeld.com
Tested to Comply
With FCC Standards
SN:
 
 
Figure 2: Rear panel of the RX10K 
 
CAUTION: The  Power Input, Voltage Output and Com (digital 
output) share a common (or ground) connection. Specifically, the 
outside ring of the Power Input, the “-“ jack of the Voltage Output 
and pin 5 of the Com output are all internally connected. 
 
Care must be taken  in connecting the RX10K to external 
components in order to eliminate ground loop problems. Such 
ground loop currents can cause errors in the output signals and 
possible internal damage. 
 
The included power supply is electrically isolated and will eliminate 
any ground loop problems with the Power Input. It is 
recommended that the Com (digital output) and Analog Output are 
not connected to external equipment at the same time. An 
exception to this rule exists when one of the two external devices 
accepting the  analog or digital output signal is “floating” or not 
externally connected, such as battery operated voltmeter or a 
laptop powered by batteries.