Inspection and Setup
Inspection and Setup
Inspect the HF PARTNER device, printer and printer adapter package contents for 
any signs of damage or defects. If any defect or damage is visible, contact your 
Guidant sales representative or call Guidant Technical Services. See “Contacting 
Guidant” on page 5.
The following steps explain how to setup your HF PARTNER system:
1.
Connect the USB cable to the printer adapter and the USB port on the printer.
2.
Attach the printer adapter power cord to the printer adapter and plug the printer 
adapter power cord into a standard wall outlet.
3.
Attach the printer power cord to the printer power input and plug the other end of 
the power cord into a standard wall outlet.
4.
Press the printer ON button.
Press the green INTERROGATE button 
 to ensure that the HF PARTNER device 
is operational; the symbols on the front will light up and a voice message will be 
heard.
How to Interrogate an Implanted Heart Failure 
Device
Use the following steps to interrogate an implanted heart failure device using the 
HF PARTNER device:
1.
Press the green INTERROGATE button 
. The “Interrogation requested” voice 
message will be heard, followed by “Locate implanted device”.
2.
Move the HF PARTNER device over the patient’s implanted device (Figure 3). 
The visual indicators on the front of the HF PARTNER device will light in 
consecutive sequence until the communication with the implanted device is 
established. During interrogation, the Interrogation light 
 will blink while data 
is transmitting. At the end of a successful interrogation, the HF PARTNER device