Q-TECH™ PROGRAMMER: OPERATiOn
Figure 47: Capture all sense vectors
Once captured, the three S-ECGs can be viewed by following the steps outlined in “To view a
Beeper Control
The SQ-RX device has an internal alert system that emits an audible beeping tone when an alert
condition occurs. Alert conditions may include Elective Replacement Indicator (ERI), End of Life
(EOL), electrode impedance out of range, prolonged charge times, failed Device Integrity Check and
irregular battery depletion. Beeper Control allows activation or suppression of the beeper.
Once an alert condition is triggered, audible tones beep for 16 seconds every nine hours until the
alert condition is resolved. Beeper Control is automatically enabled when the SQ-RX device is taken
out of Shelf mode.
To reset the Beeper Control:
Note: Access to the Beeper Control display screen is enabled only when an alert
condition occurs. If an alert condition is activated, a notification screen
will appear upon connection.
1.
From the Utilities screen (accessible from the Main Menu screen), select the
Beeper Control button.