Data Critical Corporation 02DT-7000 사용자 설명서

다운로드
페이지 62
TROUBLESHOOTING
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
 
219
All information contained herein is subject to the rights and restrictions on the title page.
spikes and display them in very short, regular intervals that are superimposed on the 
patient’s waveform. The actual pacemaker pulse voltage is typically a minimum of ten 
times the impedance pulse voltage. Repositioning the electrodes a minimum of three 
to four inches from the pacemaker will typically minimize the likelihood of the 
impedance pulse being deleted as a pacemaker spike.
Problem
False Pacer Spikes
The appearance of random, and sometimes frequent, pacemaker spikes may occur in 
some of the patients’ ECG strips.
Fig. 110. Example 13 - False Pacer Spikes
Correction
1. Turn the pacer process off if the patient has no pacemaker rhythm.
2. Always follow the recommended skin preparation procedure, as much of the 
false pacemaker activity originates from patients where poor preparation is 
present.
3. Make sure that all lead wires are connected to the electrodes securely. If you 
are intentionally leaving lead wires disconnected so that you only have to 
place three electrodes on the patient, then change to a three lead wire set.
4. Position electrodes at least three to four inches away from the pacer implant 
area.