Operator’s Manual ZM-520PA/521PA/530PA/531PA
7
Rear Panel
Refer to the “Explanations
of the Symbols in this
Manual and Instrument”
section.
of the Symbols in this
Manual and Instrument”
section.
Refer to the WARNING
below.
below.
Refer to the WARNING
below.
below.
Refer to the “Explanations
of the Symbols in this
Manual and Instrument”
section.
of the Symbols in this
Manual and Instrument”
section.
ZM-530PA/531PA only
Battery case:
Contains two alkaline dry
cell batteries (AA TYPE).
Contains two alkaline dry
cell batteries (AA TYPE).
WARNING
Close the battery case cover during operation. If the transmitter is used with the
battery case cover open, anyone who touches the opened battery case may receive
an electrical shock when defibrillation is performed. Touching the opened battery
case may cause electrostatic discharge and intermittently interfere with the waveform
or data.
battery case cover open, anyone who touches the opened battery case may receive
an electrical shock when defibrillation is performed. Touching the opened battery
case may cause electrostatic discharge and intermittently interfere with the waveform
or data.
WARNING
This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop
using it and contact your Nihon Kohden representative. If a wet transmitter is used,
the patient or operator may receive an electrical shock or injury.
using it and contact your Nihon Kohden representative. If a wet transmitter is used,
the patient or operator may receive an electrical shock or injury.