Nihon Kohden Corporation ZM-531PA 사용자 설명서
42
Operator’s Manual ZM-520PA/521PA/530PA/531PA
Electrode Positions for Respiration Monitoring
Place the R/RA and F/LL electrodes so that the lungs are between the electrodes.
Place the R/RA and F/LL electrodes so that the lungs are between the electrodes.
Position 1
In this position, respiration measurement is available; however, there is a difference in amplitude
between different patients.
In this position, respiration measurement is available; however, there is a difference in amplitude
between different patients.
R or RA
F or LL
Right infraclavicular fossa
Fifth intercostal space on the
left midclavicular line, V4
left midclavicular line, V4
Position 2
In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII.
This position can be generally recommended.
In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII.
This position can be generally recommended.
R or RA
F or LL
Right infraclavicular fossa
Fifth intercostal space on the
left midaxillary line, V6
left midaxillary line, V6
Position 3
In this position, the respiration waveform is optimum, but the ECG lead is unusual.
In this position, the respiration waveform is optimum, but the ECG lead is unusual.
R or RA
F or LL
Right midaxillary at the
horizontal level of V6
horizontal level of V6
Fifth intercostal space on the
left midaxillary line, V6
left midaxillary line, V6
R/RA
(−)
(−)
F/LL
(+)
(+)
N
R/RA
(−)
(−)
F/LL
(+)
(+)
N
R/RA
(−)
(−)
F/LL
(+)
(+)
N
R/RA
(−)
(−)
F/LL
(+)
(+)
N
N
R/RA
(−)
(−)
F/LL
(+)
(+)
N
R/RA
(−)
(−)
F/LL
(+)
(+)