ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
The bioelectric impedance measurement sensor of 
a minute ventilation rate-adaptive implantable 
pacemaker may be affected by transmitter which is 
connected to the same patient. If this occurs, the 
pacemaker may pace at its maximum rate and the 
transmitter may give incorrect data to the monitor. 
If this occurs, disconnect the electrode leads from 
the patient or change the setting on the pacemaker 
by referring to the pacemaker’s manual. For more 
details, contact your pacemaker representative or 
Nihon Kohden representative.
WARNING
Turn the pacing pulse detection to ON on the 
receiving monitor when monitoring a pacemaker 
patient. Otherwise the pacemaker pulse is not 
rejected. However, even when the pacing pulse 
detection is set to ON, the pacemaker pulse might 
not be rejected. When the pacemaker pulse is not 
rejected, the pacemaker pulse is detected as QRS 
and false heart rate may be indicated or critical 
arrhythmia such as asystole may be overlooked. 
Keep pacemaker patients under close observation.
)RUWKHSDFHPDNHUSXOVHUHMHFWLRQFDSDELOLW\RIWKH=03$DQG=03$WUDQVPLWWHUV
UHIHUWRWKH³6SHFL¿FDWLRQV(&*´VHFWLRQ
WARNING
Even when the pacing pulse detection is set to ON 
on the receiving monitor, the pacemaker pulse can 
be overlooked or detected as QRS. You cannot 
confirm the pacemaker operation only from the 
detected pacemaker pulse.