Nihon Kohden Corporation ZM-541PAA 사용자 설명서
90
ZM-540PA, ZM-541PA Operator’s Manual
WARNING
The bioelectric impedance measurement sensor of
a minute ventilation rate-adaptive implantable
pacemaker may be affected by transmitter which is
connected to the same patient. If this occurs, the
pacemaker may pace at its maximum rate and the
transmitter may give incorrect data to the monitor.
If this occurs, disconnect the electrode leads from
the patient or change the setting on the pacemaker
by referring to the pacemaker’s manual. For more
details, contact your pacemaker representative or
Nihon Kohden representative.
a minute ventilation rate-adaptive implantable
pacemaker may be affected by transmitter which is
connected to the same patient. If this occurs, the
pacemaker may pace at its maximum rate and the
transmitter may give incorrect data to the monitor.
If this occurs, disconnect the electrode leads from
the patient or change the setting on the pacemaker
by referring to the pacemaker’s manual. For more
details, contact your pacemaker representative or
Nihon Kohden representative.
WARNING
Turn the pacing pulse detection to ON on the
receiving monitor when monitoring a pacemaker
patient. Otherwise the pacemaker pulse is not
rejected. However, even when the pacing pulse
detection is set to ON, the pacemaker pulse might
not be rejected. When the pacemaker pulse is not
rejected, the pacemaker pulse is detected as QRS
and false heart rate may be indicated or critical
arrhythmia such as asystole may be overlooked.
Keep pacemaker patients under close observation.
receiving monitor when monitoring a pacemaker
patient. Otherwise the pacemaker pulse is not
rejected. However, even when the pacing pulse
detection is set to ON, the pacemaker pulse might
not be rejected. When the pacemaker pulse is not
rejected, the pacemaker pulse is detected as QRS
and false heart rate may be indicated or critical
arrhythmia such as asystole may be overlooked.
Keep pacemaker patients under close observation.
)RUWKHSDFHPDNHUSXOVHUHMHFWLRQFDSDELOLW\RIWKH=03$DQG=03$WUDQVPLWWHUV
UHIHUWRWKH³6SHFL¿FDWLRQV(&*´VHFWLRQ
UHIHUWRWKH³6SHFL¿FDWLRQV(&*´VHFWLRQ
WARNING
Even when the pacing pulse detection is set to ON
on the receiving monitor, the pacemaker pulse can
be overlooked or detected as QRS. You cannot
confirm the pacemaker operation only from the
detected pacemaker pulse.
on the receiving monitor, the pacemaker pulse can
be overlooked or detected as QRS. You cannot
confirm the pacemaker operation only from the
detected pacemaker pulse.