ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
When performing NIBP measurements in STAT 
mode or 5 minute intervals, periodically remove the 
cuff from the patient for ventilation. The skin 
temperature may increase at the cuff attachment 
site by 2 or 3°C (4 or 5°F). When measuring a 
patient with a fever or peripheral circulation 
insufficiency, it may cause a burn.
WARNING
When performing long term measurement at 
intervals less than 5 minutes, perform 
measurements while observing the state of the 
patient, blood vessels and limb to ensure adequate 
circulation. Congestion may occur at the 
measurement site. When performing periodic 
measurement for a long time, periodically check 
the circulation condition.
When performing NIBP measurement repeatedly, 
have a rest between measurements to recover 
adequate circulation.
CAUTION
NOTE  •  When measuring patients who are conscious, help the patient to relax. 
Measurement may not be accurate if the patient’s arm is tense or if the 
patient talks.
•  The data for measurement on a leg tends to be higher than measurement 
on the arm. When making diagnosis based on the NIBP values, measure 
NIBP on an upper arm.
•  Do not apply pressure to the cuff or air hose. NIBP may not be measured 
correctly because of noise or NIBP measurement may stop due to the 
NIBP safety circuit.
•  When the transmitter is attached to the patient arm and the NIBP 
measurement is performed when moving, tell the patient to relax and keep 
quiet. Otherwise, measurement may be stopped or remeasurement is 
repeated due to body movement.
•  If there is an abnormal noise generated during measurement, stop using 
the transmitter and contact your Nihon Kohden representative.