ZM-540PA, ZM-541PA  Operator’s Manual
Changing the Transmitter Channel
7KHFKDQQHORIWKHWUDQVPLWWHUFDQEHFKDQJHGZLWKDQRSWLRQDO4,3.FKDQQHOZULWHU
WARNING
The following actions must be taken to properly 
receive the transmitter signal of the correct patient 
on the receiving instrument. Otherwise, there may 
be signal loss or signals may mix causing a serious 
accident, such as monitoring a different patient.
•  Assign a channel administrator in the hospital 
and only he or she should manage channel 
assignment.
•  The channel administrator must manage the 
channels in the facility so that there is no signal 
interference.
•  When the transmitter channel is changed, the 
channel administrator must check that the 
channel on the receiving monitor is also 
changed and the signal is properly received.
•  The channel administrator must replace the 
channel number label on the transmitter with the 
new one after changing the channel.
NOTE  •  The software version of the QI-901PK channel writer must be 02-01 or later 
to change the channel on the transmitter.
•  The channel writer must be used outside the patient environment.
7KHFKDQQHOLVGLVSOD\HGLQWKHXSSHUOHIWFRUQHURIWKHVFUHHQ
Channel